<%@ Page Language="C#" AutoEventWireup="True" Debug="true" %> UDCO 8B6 Calibration Certificate
- trusted supplier since 1959
Certificate of Calibration

Company Logo

UNHOLTZ-DICKIE CORPORATION

Vibration Testing Equipment

6 Brookside Drive

Wallingford, CT-06492

Tel: (203) 265-3929

Fax: (203) 265-2690


Certificate of Calibration

For Accelerometer in compliance with MIL-STD-45662A


  Customer: XYZ Corp

  Customer P.O: RQPNM


Transducer Manufacturer: U-D

Model: 8B6

S/N: 410

Ambient temperature: 75°F±15°F

Relative humidity: <85%



CALIBRATION RESULTS:


REF. LEVEL: 2.00 g

REF. FREQUENCY: 160.0 Hz

SENSITIVITY: 7.987 pC/g


Freq Sensitivity Deviation
(Hz) (pC/g) (%)

10.00

15.00

30.00

50.00

100.00

160.00

200.00

300.00

400.00

500.00

600.00

700.00

7.97

7.98

7.98

7.98

7.98

7.99

7.99

7.99

7.99

7.99

7.99

7.99

-0.18

-0.11

-0.04

-0.04

-0.03

0.00

0.02

0.01

0.02

0.04

0.05

0.06

Freq Sensitivity Deviation
(Hz) (pC/g) (%)

800.00

900.00

1000.00

1500.00

1700.00

2000.00

2500.00

3000.00

3500.00

4000.00

4500.00

5000.00

7.99

7.99

7.99

8.00

8.01

8.02

8.04

8.06

8.08

8.11

8.15

8.17

0.07

0.08

0.10

0.22

0.28

0.38

0.62

0.86

1.18

1.57

2.00

2.35

Freq Sensitivity Deviation
(Hz) (pC/g) (%)

5500.00

6000.00

6500.00

7000.00

7500.00

8000.00

8500.00

9000.00

9500.00

10000.00

*

*

8.22

8.27

8.32

8.38

8.44

8.51

8.59

8.66

8.76

8.85

2.97

3.50

4.22

4.90

5.70

6.58

7.51

8.48

9.62

10.82

Sensitivity-Deviation Vs Frequency


Model 8B6 SN 410


Calibrated by:


Calibration Date: 02/22/2006

(Operator's Name)



ESTIMATED UNCERTAINTY: 10Hz to 2kHz; +/- 2.1% from 2kHz to 4kHz; +/- 2.4% at 4 kHz increasing to +/-3.3% at 10kHz.

TRACEABILITY: This calibration has been obtained by using Accelerometer measuring and Reference Standards whose calibration is scheduled at adequate intervals to assure the stated accuracy and is traceable to NIST Test No. 822/263866 19-Jul-2005. Reference Accel due 19-Jul-2006.

EQUIPMENT USED FOR CALIBRATION: UDVwin Transducer Calibration System with,

Lab Standard Transducer: 8B6

Serial Number: 345

Signal Conditioner Model: TSC-3

Serial Number: 108

Variable Gain Amplifier Board Model: VGN-8

Serial Number: 427

Digital Signal Processor Board Model: DSP-8

Serial Number: 00100100

Working STD Calibration: std2001-run1 on 01/22/2006